Exper,mental study and clinical observation of minimumcontact plate in long bone fracture

在线阅读 下载PDF 导出详情
摘要 Tosutdythemechanicalpropertiesandtheclinicalresultsofminimum-contactplateinthetreatmentoffractures.Methods:Four-pointbendingandtorsiontestswereconductedtocomparethenewminimumcontactplate(MCP)withdynamiccompressionplate(DCP)andlimitedcontactdynamiccompressionplate(LC-DCP).Thefracturehealingtimeandgrowthofhonycalluswereobservedin29casesoflongbonefracturefixedwithMCPtoevaluatetheadvantagesofthiskindofplate.Results:The29patientswhounderwentMCPwerefollowedupfrom6to14months.Theaveragehealingtimeforfemoral,tibialandhumeralfractureswas12,13and10weeks,respectively.Allfractureswerehealed.Nodisplacementoffracture,screwpullout,deformationorhreakoftheplatewerefound.Conclusions:ThebendingstiffnessofMCPissignificantlygreaterthanthatofDCPandLC-DCP(P<0.05).MCPcanprotecttheperiostealhloodsupplyagainstavascularosseousnecrosisandacceleratehonehealing.Itisakindofreliableandeffectiveplateintreatmentoffractures.
机构地区 不详
出版日期 2005年02月12日(中国期刊网平台首次上网日期,不代表论文的发表时间)